أنشئ حسابًا أو سجّل الدخول للانضمام إلى مجتمعك المهني.
Generally TCD is used to measure Percentage levels, however the MDL can vary between manufactures. The PPM measurement through TCD is acceptable since TCD is a very Linear Detector and the key performer in TCD PPM detection is the Sample Load. The minimum a good TCD Can measure, which i have witnessed is 10~15 ppm.
TCD is used to measure the Percentage level, However the minimum level of accuracy depends to its noise ratio
TCD is can be used to measure upto 1-10 ppm for gases such as Argon or Oxygen, CH4, Carbon monoxide, Hydrogen, Nitrogen etc.
FID is used for measuring Hydrocarbons and usually can have range from 1-10 ppm as well.
TCD are used to measure Higher concentrations of hydro carbons in %. FPD and FID are used in PPM level depends upon the process.