Yes. Since the publication of the inspection guide on cleaning validation in 1993, a number of studies have been published to demonstrate the adequacy of TOC in measuring contaminant residues.
We think TOC can be an acceptable method for monitoring residues routinely and for cleaning validation. But in order for TOC to be functionally suitable, it should first be established that a substantial amount of the contaminating material is organic and contains carbon that can be oxidized under TOC test conditions. This is not a trivial exercise because we know that some organic compounds cannot be reliably detected using TOC.
TOC use may be justified for direct surface sample testing as well as indirect (rinse water) sample testing. In either case, because TOC does not identify or distinguish among different compounds containing oxidizable carbon, any detected carbon is to be attributed to the target compound(s) for comparing with the established limit. Thus, a firm should limit 'background' carbon (i.e., carbon from sources other than the contaminant being removed) as much as possible.
If TOC samples are being held for long periods of time before analysis, a firm should verify the impact of sample holding time on accuracy and limit of quantitation.