Inscrivez-vous ou connectez-vous pour rejoindre votre communauté professionnelle.
You regenerate HPLC columns by flushing it with the following volumes of solvent.
by multiple washing with mobile phase they can be regenerate
HPLC columns is regenerate by washing washing with purified water and preserve with methanol or Acetonytrle 50% depend open the type of mobile phase we are using
To clean and regenerate a reversed phase column flush it with the following volumes of solvent.
BY PASSING 1 M NACL SOLUTION THROUGH CHROMATOGRAPHY COLUMN.
For HPLC Reverse phase;
wash with unbuffered mobile phase
wash with100% Water
wash with Methaol or ACN
wash with THF or IPA
wash with Methylene chloride
wash with N-Heptane
wash with Methylene chloride
wash with Methanol or ACN
Wash with Water
Return to Solvent
Time:10 times of column volume.
Column volume: Πr2h
Example: column dimension250*4.6*5
h=250mm=25cm
r=4.6/2 mm=0.23cm
Π=3.14
Volume=3.14*0.23*0.23*25 =4.1527cm3=4.1527 ml
Flow=1.0ml/min
Time: (10*Vol)/flow =10*4.1527/1.0=41.52 minutes
42 minutes enough for each washing.
regeneration done by organic solvent like acetonitril
For Ion Chromatographic columns, you can regenerate your column by one the following methods:-
1. Hydrophilic contamination of low valence: Concentrated eluent of10 times the concentration of the most concentrated eluent used in the application.
2. High valence hydrophilic ions contamination & Metal Contamination:1 to3 M HCl.......if Metal Contamination still there try with0.2 Molar Oxalic acid.
3. Nonionic and hydrophobic contamination:80% Acetonitrile solution will be used.
For HPLC columns; Flush it with water in the reverse direction at a flow of1ml/min. for about30 minutes, then flow Isopropanol, Methylene Chloride, Hexane, Methylene Chloride, then Isopropanol at this exact order each for30 minutes at a flow of1 ml/min.
Then flow the mobile phase you're working with at the normal direction of the column but without the buffer for30 minutes then introduce the buffer.
After this, performance should be restored for the column.