How the HPLC colum and ion chromatgraphic colum is re genrate

user-image
Question ajoutée par Syed Ashraf , Chemist , Merck Sharp & Dohem
Date de publication: 2014/10/28
Aavian Hapanyengwi
par Aavian Hapanyengwi , qc chemist , Kiarah Chemicals CC

You regenerate HPLC columns by flushing  it with the following volumes of solvent.

  • 20 column volumes of water/acetonitrile (95:5 v/v)
  • 20 column volumes of acetonitrile
  • 5 column volumes isopropanol
  • 20 column volumes hexane
  • 5 column volumes isopropanol
  • 20 column volumes acetonitrile
  • 20 column volumes of water/acetonitrile (95:5 v/v)
  • Re-equilibrate with the required mobile phase                                                                                                     Ion chromatographic columns are regenerated by flushing with electrolytes such as acids and bases                                                    

Zisan  Ahamad
par Zisan Ahamad , Project Trainee , CSIR-CDRI

by multiple washing with mobile phase they can be regenerate

Faraz ASKARI ABIDI
par Faraz ASKARI ABIDI , Quality control officer , Pfizer Pakistan

HPLC columns is regenerate by washing washing with purified water and preserve with methanol or Acetonytrle 50% depend open the type of mobile phase we are using

Virendra Vikhare
par Virendra Vikhare , chemist , Rezilabs for Industrial Services / SAECO

To clean and regenerate a reversed phase column flush it with the following volumes of solvent.

  •  water/acetonitrile (95:5 v/v)
  •  acetonitrile
  •  isopropanol
  • hexane
  •  isopropanol
  •  acetonitrile
  • water/acetonitrile (95:5 v/v)
  • Re-equilibrate with the required mobile phase

ABDULRAHIM MAHEBOOB MOMIN MOMIN
par ABDULRAHIM MAHEBOOB MOMIN MOMIN , SR OFFICER , WOCKHARDT BIOTECH PARK

BY PASSING 1 M NACL SOLUTION THROUGH CHROMATOGRAPHY COLUMN.

KRISHNAMOORTHI THANGAVEL
par KRISHNAMOORTHI THANGAVEL , Research Associate , Shell technology center, Bangalore, India

For HPLC Reverse phase;

wash with unbuffered mobile phase

wash with100% Water

wash with Methaol or ACN

wash with THF or IPA

wash with Methylene chloride

wash with N-Heptane

wash with Methylene chloride

wash with Methanol or ACN

Wash with Water 

Return to Solvent

Time:10 times of column volume.

Column volume:   Πr2h

Example: column dimension250*4.6*5

h=250mm=25cm

r=4.6/2 mm=0.23cm

Π=3.14

Volume=3.14*0.23*0.23*25 =4.1527cm3=4.1527 ml

Flow=1.0ml/min

Time: (10*Vol)/flow =10*4.1527/1.0=41.52 minutes 

42 minutes enough for each washing.

Bishoy Ashraf
par Bishoy Ashraf , Quality control senior , Global NAPI Pharmaceutical

regeneration done by organic solvent like acetonitril

Ashraf Mohamed
par Ashraf Mohamed , Senior Laboratory Chemist , Arab Company for Laboratories and soil

For Ion Chromatographic columns, you can regenerate your column by one the following methods:-

1. Hydrophilic contamination of low valence: Concentrated eluent of10 times the concentration of the most concentrated eluent used in the application.

2. High valence hydrophilic  ions contamination & Metal Contamination:1 to3 M HCl.......if Metal Contamination still there try with0.2 Molar Oxalic acid.

3. Nonionic and hydrophobic contamination:80% Acetonitrile solution will be used. 

Wessam AbuShousha
par Wessam AbuShousha , Analyst (Chemist) , Pharco Pharmaceuticals

For HPLC columns; Flush it with water in the reverse direction at a flow of1ml/min. for about30 minutes, then flow Isopropanol, Methylene Chloride, Hexane, Methylene Chloride, then Isopropanol at this exact order each for30 minutes at a flow of1 ml/min.

Then flow the mobile phase you're working with at the normal direction of the column but without the buffer for30 minutes then introduce the buffer.

After this, performance should be restored for the column.